In plain English
Every measurement is a little uncertain. If you measure the same shaft several times you get slightly different numbers, the instrument has its own small errors, and temperature changes the part. Measurement uncertainty puts a number on that doubt, so a result can be written as, for example, 25.012 mm ± 0.003 mm.
Technical definition
Measurement uncertainty is a non-negative parameter characterising the dispersion of the values attributed to a measured quantity, based on the information used (VIM). A standard uncertainty is expressed as a standard deviation. The combined standard uncertainty uc combines all the contributions. The expanded uncertainty U = k × uc defines an interval expected to contain a large fraction of the values, where k is the coverage factor.
How it works and where it is used
The GUM method builds an uncertainty budget:
- List every significant source — calibration of the instrument, resolution, repeatability, temperature, the operator, the part’s form.
- Evaluate each as a standard uncertainty: Type A from statistics of repeated readings, Type B from other information such as certificates, specifications or resolution.
- Combine them (by root-sum-square when independent), then multiply by k.
Formula
uc = √[ Σ (ci × ui)² ] U = k × uc
Rectangular limits ±a: u = a ÷ √3 From a certificate: u = U ÷ k
ui = standard uncertainty of each independent component, ci = its sensitivity coefficient. k = 2 gives roughly 95% coverage when the result is close to normally distributed with enough degrees of freedom.
Worked example
Measuring a length: certificate U = 1.5 µm (k = 2) → u = 0.75 µm; display resolution 1 µm → ±0.5 µm rectangular → u = 0.29 µm; repeatability u = 0.8 µm.
uc = √(0.75² + 0.29² + 0.8²) = 1.13 µm; U = 2 × 1.13 = 2.3 µm. The result is reported as, for example, 25.0120 mm ± 0.0023 mm (k = 2).
In the workshop and in practice
Uncertainty decides whether a measurement can prove a part is in tolerance. ISO 14253-1 sets the default rule: to prove conformance, the measured value must lie inside the tolerance zone reduced by the expanded uncertainty at each end. If U is a large share of the tolerance, good parts get rejected or bad parts accepted — the instrument or method has to improve. Calibration laboratories state uncertainty on every certificate.
Common mistakes and misconceptions
- Confusing uncertainty with error. Error is the (unknown) difference from the true value; uncertainty describes the doubt about the result.
- Using resolution alone. Resolution is only one component; repeatability and calibration are usually larger.
- Adding uncertainties directly. Independent standard uncertainties combine by root-sum-square, not by simple addition.
- Not stating k. “± 2.3 µm” without the coverage factor is incomplete.
Related calculators
- Measurement Uncertainty Calculator (GUM) — Combined standard and expanded uncertainty from independent components (root-sum-square).
- Vernier Caliper Least Count & Reading Calculator — Least count of a vernier scale, and the reading with zero-error correction.
- Micrometer Least Count & Reading Calculator — Least count from pitch and thimble divisions, and a micrometer reading with zero error.
Related topics
- Least Count — The smallest change in size that a measuring instrument can show.
- Limits of Size and Tolerance — The largest and smallest acceptable sizes of a feature, and the tolerance between them.
References
- JCGM 100:2008 Evaluation of measurement data — Guide to the expression of uncertainty in measurement (GUM)
- JCGM 200:2012 International vocabulary of metrology (VIM)
- ISO 14253-1: GPS — Inspection by measurement of workpieces and measuring equipment — Decision rules for verifying conformity or nonconformity with specifications